6.2 Alarm Modes
When an alarm occurs, the monitor raises the user’s attention by the following audible or
visual indications.
Visual alarms
Audible alarms
Alarm messages
Parameter flashes
Besides, the visual alarms, audible alarms and alarm messages are given in different ways to
identify different alarm levels.
6.2.1 Visual Alarms
The alarm indicator on the front panel of the monitor varies its flash color and frequency to
indicate different alarm levels.
High level alarm: red and quick flash;
Medium level alarm: yellow and slow flash;
Low level alarm: yellow light without flash.
6.2.2 Audible alarms
The monitor uses different alarm tones to indicate different alarm levels.
High level alarm: “DO-DO-DO--DO-DO---DO-DO-DO--DO-DO”.
Medium level alarm: “DO-DO-DO”.
Low level alarm: “DO”.
Different intervals correspond to different alarm levels: High level alarm phonates once
every 3 or 8 seconds. Medium level alarm phonates once every 14 or 24 seconds. Low level
alarm phonates once every 24 seconds. For details, please refer to 4.7 Maintenance.
NOTE
When multiple alarms of different levels occur simultaneously, the monitor
selects the alarm of the highest level and gives alarm tone accordingly.
6-3