Alarm message A B L Cause Measure
RECORDER
PAPER W.P.
Yes Yes 2 The paper roll of the
recorder is not placed in
the correct position.
Place the paper roll correctly.
RECORDER S.
COMM ERR
Yes No 2 Error in recorder
communication.
Open the RECORD menu
and select the CLEAR REC
TASK option. If the problem
remains, contact our
company for repair.
REC NOT
AVAILABLE
No No 2 Error in the recorder work
mode.
C.2.13 System Alarm Messages
Alarm message A B L Cause Measure
REAL CLOCK
NEED SET
No No 1 The system time is
incorrect.
Reset the system time and
then restart the monitor.
REAL CLOCK
NOT EXIST
No No 1 No button battery, or the
battery power is depleted.
Add, or replace with a new
button battery.
KEYBOARD
INIT ERR N
No No 1 Keyboard error. The
keyboard cannot be used.
Contact our company for
repair.
Note: N represents the error
number.
KEYBOARD
ERROR
No No 2
NET INIT ERR
(G.)
No No 2 The system cannot be
connected to the network