In this menu, you can perform the following settings:
ST ANAL ST analysis
ON: Enables the ST analysis;
OFF: Disables the ST analysis.
NOTE
When turning ST ANALYSIS on, the monitor selects DIAGNOSTIC mode
automatically. You can set the monitor to MONITOR or SURGERY mode as
required. However, the ST numerics might be severely distorted in these modes.
ST ALM ST segment alarm
ON: If the measured ST numerics exceed the alarm limit,
the monitor gives alarm indications and saves the alarm;
OFF: If the measured ST numerics exceed the alarm limit,
the monitor does not give alarm indications or saves the
alarm.
When OFF is selected, the
icon is displayed on the
right of ST1 in the parameter window.
ST ALM ST alarm level
Options: HIGH, MED and LOW;
ALM REC ST alarm recording
ON: The monitor starts recording when an ST alarm occurs;
OFF: The monitor does not record when an ST alarm
occurs.
ALM HI Determines the upper limit of the ST alarm; 2.0mV is the
highest.
ALM LO Determines the lower limit of the ST alarm; -2.0 mV is the
lowest.
NOTE
The alarm limits for two ST segment numerics are identical. You cannot set the
alarm limit of one channel separately.
11-16