In this menu, you can select any of the displayed option. All options are listed below.
SMART LEAD
OFF
ON: If there is a LEAD OFF in the HR-derived channel, the
system will automatically switch to another lead to restore the
display of the ECG waveform in the HR-derived channel.
OFF: The SMART LEAD OFF is turned off.
When ECG is in NORMAL DISPLAY mode, if the SMART LEAD OFF is configured ON
and there is a LEAD OFF in the HR-derived channel, the system will automatically switch to
another lead to restore the display of the ECG waveform in the HR-derived channel. When
the LEAD OFF condition is corrected, the leads are automatically switched back. However, if
you manually switch the HR-derived channel, change the ECG display mode, or turn off
SMART LEAD OFF before the LEAD OFF condition is corrected, the leads will not be
switched back.
Switching rules for SMART LEAD OFF:
LEAD OFF condition HR-derived channel automatically switches to
RA OFF III
LL OFF I
LA OFF II
V OFF II
RL OFF None
Any two out of LL, RA and LA fall off. None
LL, RA or LA, and V fall off together Same as LL, RA or LA falls off.
NOTE
The SMART LEAD OFF function is available only when the LEAD TYPE is 5
LEADS.
When the HR CHANNEL is set to AUTO, SMART LEAD OFF does not work as
long as there is ECG waveform displayed in either channel 1 or 2.
When you perform ECG calibration, SMART LEAD OFF does not work.
11-11