Shenzhen PM-9000 Express Manual del operador Pagina 201

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 339
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 200
Temperature alarm limits in factory configuration:
Parameter Patient type Upper limit Lower limit
Feline 39.5 37.5
T1 and T2 Canine 39.5 38
Other animals 39.5 37.5
Feline 2 /
TD Canine 2 /
Other animals 2 /
14.4 Maintenance and Cleaning
WARNING
Before cleaning the monitor or the probe, make sure the equipment is turned off
and disconnected from AC power.
Reusable Temperature Probes
The temperature probe should not be heated to a temperature over 100. (212.). It is
only able to bear the temperature between 80 and 100. (176 to 212.) for a short time.
The probe must not be disinfected in steam.
Only detergents containing alcohol can be used for disinfection.
The rectal probes should be used, if possible, in conjunction with a protective rubber
cover.
To clean the probe, hold the tip with one hand and with the other hand rub the probe
down in the direction of the connector using a moist lint-free cloth.
NOTE
Disposable temperature probes must not be re-sterilized or reused.
Disposable temperature probes must be recycled or disposed of properly.
14-4
Vista de pagina 200
1 2 ... 196 197 198 199 200 201 202 203 204 205 206 ... 338 339

Comentarios a estos manuales

Sin comentarios