3 Installation
WARNING
•
The installation of the monitor must be carried out by personnel authorized by us.
The software copyright of the monitor is solely owned by our company. Any action
to change, copy or exchange the software copyright by any organization or person
is regarded as copyright infringement and is not allowed.
•
Accessory equipments connected to this patient monitor must be certified
according to the respective IEC standards (e.g. IEC 60601-1 for medical electrical
equipment). Furthermore all configurations shall comply with the valid version of
the system standard IEC 60601-1-1. Any person who connects additional
equipment to the signal input or signal output is responsible to ensure the system
complies with the requirements of the valid version of the system standard IEC
60601-1-1. If in doubt, contact our company or customer service.
•
If the monitor is connected to another electrical instrument and the instrument
specifications cannot tell whether the instrument combination is hazardous (e.g.
due to summation of leakage currents), you should consult us or experts in the field
to ensure the required safety of all instruments concerned.
NOTE
•
The operations in this section are not all required. User-customized installation by
authorized personnel is provided.