Pressure transducer
interface cable
Monitor
Pressure transducer
interface cable
Monitor
15.4.2 IBP Pressure Zero Menu
You can select IBP PRESURE ZERO in IBP (1,2) SELECT menu to open the following
menu.
Figure 15-9 IBP Pressure Zero Menu
Pressure Transducer Zeroing Procedure
1.
Disconnect the transducer from the patient.
Pressure
transducer
3-way
stopcock
Figure 15-10 Pressure Transducer Zeroing
2.
Adjust the 3-way stopcock to close the channel leading to the patient. The transducer is
open to the atmosphere through the stopcock.
3.
Select CH1 ZERO or CH2 ZERO in IBP PRESSURE ZERO menu to start zeroing.
15-8