Alarm message A B L Cause Measure
OVER
PRESSURE
No Yes 2 The airway might be blocked. proper cuff and connect
it correctly. If the
problem still exists, NIBP SYSTEM No Yes 2 Failure occurs in the pulse
FAILURE measurement. The monitor
cannot perform measurement,
analysis, or calculation.
contact our company for
repair.NIBP TIME
OUT
No Yes 2
MEASURE
FAIL
No Yes 2
NIBP
ILLEGALLY
No Yes 2 Illegal reset comes out during
the NIBP measurement.
Check if the airway is
blocked. Deal with the
RESET blocking and perform the
measurement again. If
the problem still exists,
contact our company for
repair.
C.2.6 Mindray SpO2 Module Alarm Messages
Alarm message A B L Cause Measure
SPO2 SENSOR No Yes 3 The sensor is disconnected Make sure the sensor is
OFF from the patient or the
monitor.
placed on the patient’s
finger or other parts, and
the monitor is connected to
cables correctly.
SPO2 NO Yes Yes 3 The sensor is disconnected Disconnect and reconnect
SENSOR from the patient or the
monitor, or it is not
the sensor as directed by
the instructions. If the
properly connected. alarm remains, the sensor
or the cable might have
been damaged.
The SpO2 is connected
upside down.
Disconnect and reconnect
the sensor as directed by
the instructions. Pay
Vista de pagina 311