14 TEMP Monitoring
14.1 Overview
The monitor is able to use two temperature probes simultaneously, to obtain two temperature
values and the difference between them. The measurement values are displayed in the TEMP
parameter window as shown below.
1
4
2
3
5
Figure 14-1 TEMP Parameter Window
1.
TEMP label: Selecting this label opens the TEMP SETUP menu.
2.
Temperature unit: . or ..
3.
T1: Temperature channel 1. Displays the temperature measured at temperature channel
1.
4.
T2: Temperature channel 2. Displays the temperature measured at temperature channel
2.
5.
TD: Temperature difference, namely the difference between temperature 1 and
temperature 2.
14.2 Measurement Procedure
To measure the temperature of a patient,
1.
If a disposable temperature probe is used, plug the temperature probe cable in the
temperature probe connector on the side panel of the monitor, and then connect the
temperature probe with the cable; if a reusable temperature probe is used, connect the
temperature probe with the temperature probe connector directly.
2.
Attach the temperature probe to the patient properly.
3.
Power on the monitor.
14-1