The TIME is displayed on the left of the trend table. On the top is the latest time. From the
upper to the lower, the interval between two adjacent times depends on the preset resolution.
And the date is contained in the brackets. On the right of the TIME is the EVENT. If a
marked event happened at a specific time, the mark will be displayed aside that time in the
EVENT field. On the right of the trend table are parameter names and the trend data. The
symbol ".." means the parameter is not measured at the corresponding time. Besides, the
L-RIGHT option allows change of the parameter name and the trend data.
If you select to review the NIBP trend data, the measurement results as well as the specific
measurement time (in the TEST AT filed) are displayed. If more than one measurement
results are obtained in a time interval, only one result is to be displayed in the measurement
result field.
•
RESOLUTION
1.
Rotate the control knob to highlight the field at the right of RESOLUTION.
2.
Press the control knob, and a popup value with the options, 1MIN, 5 MIN, 10MIN,
30MIN and 60MIN, is opened.
3.
The time displayed in the TIME filed changes with the resolution.
•
UP-DOWN
1.
Rotate the control knob to highlight the UP-DOWN option, and press.
2.
If the
mark is displayed at the lower right of the TIME field, you can rotate the
control knob anticlockwise to page down and review the trend data of earlier time.
3.
If the
mark is displayed at the upper right of the TIME field, you can rotate the
control knob clockwise to page up and review the trend data of later time.
•
L-RIGHT
1.
Rotate the control knob to highlight the L-RIGHT option, and then press.
2.
Rotate the control knob to select a parameters set.
The “>” mark on the right of the parameter names indicates the following page is available,
and the “<” mark on the left of the parameter names indicates the previous page is available.
3.
The parameter names and the trend data changes with the selected parameter set.
•
REC
The REC option allows you to print out the trend data of the currently displayed
parameter(s).