UNIT
Options: mmHg, kPa;
INTERVAL Select MANUAL to set the monitor to manual NIBP
measurement mode, or select from the time options to
determine the interval between automatic measurements.
Optional intervals: 1, 2, 3, 4, 5, 10, 15, 30, 60, 90, 120, 180,
240, and 480MIN.
CUFF TYPE Options: BIG, MIDDLE, SMALL (short for big cuff, middle
cuff, small cuff)
Every CUFF TYPE corresponds to different cuff sizes, which are identified on the cuff. As
follows:
CUFF TYPE CUFF SIZE
BIG >25cm
MIDDLE 8 to 26cm
SMALL <13cm
RESET Select this option to restore the initial settings of the
pressure pump. If the monitor fails to give a visual
indication when the pressure pump is working improperly,
selecting this option activates a self-test procedure, and
restores the monitor to normal performance.
CONTINUAL Select this option to start a continuous measurement for five
minutes.
CONFIG >> You can select this option to access the NIBP CONFIG
menu, in which you may select FACTORY CONFIG or
USER CONFIG. After finishing your selection and exiting
the menu, the system pops up a dialog box asking for your
confirmation.
13-9