Shenzhen PM-9000 Express Manual del operador Pagina 287

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 339
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 286
RR
Measurement range Feline, Canine, Other animals: 0 to 150 BrPM
Resolution 1 BrPM
Precision
7 to 150 BrPM:
0 to 6 BrPM:
±2 BrPM or ±2%, whichever is greater.
Undefined.
Apnea alarm delay 10 to 40 s
A.10 SpO2 Specifications
A.10.1 Mindray SpO2 Module
SpO2
Measurement range 0 to 100%
Resolution 1%
Precision
70 to 100%:
70 to 100%:
0% to 69%:
±2 % (non-motion conditions)
±3% (in-motion conditions)*
Undefined.
Refreshing rate 1 s
PR
Measurement range 20 to 254 bpm
Resolution 1 bpm
Precision
±3 bpm (non-motion conditions)
±5 bpm (in-motion conditions)
Refreshing rate 1 s
A.10.2 Nellcor SpO2 Module
SpO2 measurement range
and precision
70 to 100%:
0% to 69%:
±2%
Undefined
PR measurement range and
precision
20 to 250 bpm: ±3 bpm
251 to 300 bpm: Undefined
Refreshing rate 1 s
A-8
Vista de pagina 286
1 2 ... 282 283 284 285 286 287 288 289 290 291 292 ... 338 339

Comentarios a estos manuales

Sin comentarios