CASCADE ECG cascade
ON: The monitor displays the waveform of each channel in
two lines.
OFF: The monitor displays the waveform of each channel in
one line.
The CASCADE can be turned ON only if the monitor is set
to NORMAL DISPLAY mode and the SCAN TYPE is set
to REFRESH.
•
DEFIB SYNC Defibrillator synchronization
ON: The function of defibrillation is enabled;
OFF: The function of defibrillation is disabled.
If the function of defibrillator synchronization is enabled and the ventricular defibrillation is
applied, 100ms/+5V defibrillator synchronization signals will be output through the auxiliary
output connector to the defibrillation equipment.
When the defibrillator synchronization is enabled, the “Defib Sync ON!” message is
displayed in the prompt information area, at the lower left corner of the screen. This message
disappears after 10 seconds. Besides, the technical alarms area also shows the “DEFIB
SYNC ON” message.
NOTE
•
Defibrillator synchronization can be enabled only if the AUX OUTPUT in the
USER MAINTAIN menu is set to DEFIB. SYN. Otherwise, this function is
disabled.
•
When the monitor is started, DEFIB SYNC is set to OFF as a default.
11-13