SYS ALM HI Determines the upper limit of the systolic pressure.
•
SYS ALM LO Determines the lower limit of the systolic pressure.
•
MEAN ALM HI Determines the upper limit of the mean pressure.
•
MEAN ALM LO Determines the lower limit of the mean pressure.
•
DIA ALM HI Determines the upper limit of the diastolic pressure.
•
DIA ALM LO Determines the lower limit of the diastolic pressure.
If a measured pressure crosses a preset upper or lower alarm limit, an alarm will be triggered.
The NIBP alarm limits are as follows:
Cuff type Big Middle Small
Systolic pressure 40–270 mmHg 40–200 mmHg 40–135 mmHg
Mean pressure 20–230 mmHg 20–165 mmHg 20–110 mmHg
Diastolic pressure 10–210 mmHg 10–150 mmHg 10–100 mmHg
The alarm limits in factory configuration (Unit: mmHg):
Cuff type Big Middle Small
Sys alarm high 150 150 110
Sys alarm low 95 85 70
Mean alarm high 116 100 90
Mean alarm low 77 55 50
Dia alarm high 100 75 80
Dia alarm low 68 40 40
•
DISPLAY WAY 1 GROUP: The NIBP parameter area only displays a group
of NIBP values obtained in the latest measurement.
GROUPS: Multiple groups of measurement values are
displayed at the lower left corner of the screen as shown
below.
NS NM ND
PR DATE TIME
108 80 70 72 2006-6-12 16:16:00
108 80 70 67 2006-6-12 16:11:00
108 84 70 68 2006-6-12 16:06:00
13-8