To ensure the accuracy of the measurement, it is suggested two consecutive measurements be
performed at an interval. The length of the interval can be set in the C.O. SETUP menu (time
unit: second). The remaining time to the next measurement is displayed at the position of 2 in
Figure 16-3. The next measurement cannot be performed until the time reduces to zero and
the “Ready for new measurement” prompt information appears.
NOTE
•
You must perform the injection within 4 seconds after pressing the START
option.
•
You are recommended to wait for 1 minute or longer, depending on the patient’s
clinical condition, before starting the next measurement.
You can perform a maximum of 6 measurements before editing. If extra measurements are
performed, the latest measurements will replace the earlier ones. If a curve in the editing
window is not selected for calculation (excluded from the averaging calculation), the place of
the curve will be taken by the new measurement result.
CO Editing
Selecting EDIT >> from WINDOW FOR C.O. MEASUREMENT opens WINDOW FOR
C.O. EDIT.
1
2
4
3
Figure 16-4 Window for CO Editing
16-6