A.13 IBP Specifications
Number of channels 2
Pressure readings Systolic, diastolic, mean pressures and PR
Pressure labels ART, PA, CVP, RAP, LAP, ICP, P1 and P2
Linear input range will be -50 to.300 mmHg, after zeroing.
Measurement range
ART 0 to 300 mmHg
PA -6 to 120 mmHg
CVP/RAP/LAP/ICP -10 to 40 mmHg
P1/P2 -50 to 300 mmHg
Resolution 1 mmHg
Precision ±2% or ±1mmHg, whichever is greater
Excitation
will be 5 Volts DC, ± 2%
Minimum load resistance will be 300O per transducer.
Update period 1s
Zero offset range ± 200 mmHg
Zero accuracy ± 1 mmHg
Noise <0.5 mmHg RTI, DC to 12.5 Hz, 300O source impedance.
Drift <0.15 mmHg/.; will not exceed ± 1 mmHg in 24 hours.
Frequency Response DC-12.5Hz ±1 Hz, -3db
PR from IBP
Measurement range 25 to 350 bpm
Precision ±1 or ±1%, whichever is greater.
Resolution 1 bpm
Pressure transducer
Sensitivity 5 uV/V/mmHg
Impedance range 300 to 3000.
A.14 CO Specifications
Measurement technique Thermal dilution
Calculated parameter CO, hemodynamics
CO 0.1 to 20L/min
Measurement range TB 23 to 43°C
TI 0 to 27°C
Resolution
CO:
TB, TI:
0.1 L /min
0.1°C
Precision
CO:
TB, TI:
±5% or ± 0.1 L /min
0.1°C
Alarm range TB: 23 to 43°C
A-10