ZERO CAL This option allows you to zero the CO2 module, so as to
eliminate the adverse affect of the baseline drift in the
process of measurement. Perform zeroing prior to
concentration calibration.
•
CONFIRM CAL This option allows you to confirm the calibration of the CO2
module.
Calibration Procedure
1.
Verify the module enters the ready-to-measure status.
2.
Connect a gas bottle with the CO2 sensor connector on the monitor as shown below.
Figure 17-6 Connections for Calibration
Monitor
Gas bottle
T-shape connector opening
to the atmosphere
Decompression
valve
Tube
3.
Fill the gas bottle with a standard gas of certain CO2 content (3%, 4%, 5%, 6% and 7%),
and input the gas to the monitor.
4.
Open the CO2 USER MAINTAIN menu, and set the CO2 field to a value, which is the
same with the CO2 content of the gas bottle.
5.
When the CO2 USER MAINTAIN menu displays the currently measured CO2 content,
wait until a stable reading is indicated, and then select the CONFIRM CAL option to
start calibrating the CO2 module.
6.
If the calibration succeeds, “CALIBRATE SUCCESS!” will be displayed in the CO2
USER MAINTAIN menu. If fails, “CAL FAIL, TRY AGAIN!” will be displayed, and
re-calibration is required.
7.
Select the EXIT option to exit the CO2 USER MAINTAIN menu.
17-10