11 ECG/RESP Monitoring
11.1 Overview
11.1.1 ECG Waveform
In the standard screen, one or two ECG waveform(s) is (are) displayed at the top of the
display when LEAD TYPE is set to 3 LEADS or 5LEADS respectively in the ECG SETUP
menu.
1 2 3
Figure 11-1 ECG Waveforms
4 5
As shown above, when 5 LEADS is selected, five labels are located above the ECG
waveforms:
1.
ECG lead of channel 1 (Primary lead)
You can select the lead of channel 1 from the label options.
When you use a 3-lead set, the monitor gives three lead options: I, II and III;
When you use a 5-lead set, the monitor gives seven lead options: II, III, aVR, aVL,
aVF, V and I.
2.
Gain of the channel-1 waveform
You may use this label to adjust the amplitude of the channel-1 ECG waveform. The
gain options include ×0.125, ×0.25, ×0.5, ×1, ×2 and AUTO. When AUTO is selected,
the monitor adjusts the gain automatically. Besides, a 1mV scale is displayed at the right
side of each ECG waveform. The height of the 1mV bar is directly proportional to the
ECG waveform amplitude.
11-1