For different patient types, the upper/lower limits of the respiration rate alarm may vary in
the following range.
Patient type Max. ALM HI Min. ALM LO Increment (breath/min)
Feline 150 0 1
Canine 150 0 1
Other animals 150 0 1
APNEA ALM Apnea alarm
Determines whether the patient’s cessation of breath is an
apnea event. Range: 10–40 seconds.
SWEEP Waveform speed
Options: 6.25, 12.5 and 25.0 mm/s.
HOLD TYPE Calculation type
Options: AUTO and MANUAL
1.
AUTO: When AUTO is selected, the monitor automatically determines the detection
threshold for respiration and calculates the respiration rate. The HOLD HI and HOLD
LO options are inactive.
2.
MANUAL: When MANUAL is selected, the user determines the detection threshold for
respiration and the monitor calculates the respiration rate depending on the user-selected
criteria.
CONFIG >> Select CONFIG >> to access the RESP CONFIG menu. You
can select either FACTORY CONF or USER CONF. After
finishing the selection and exiting the menu, a dialog box
pops up asking for confirmation of your selection.
11-28