In this menu, you can perform the following settings:
ARR ANAL Arrhythmia analysis
ON: Enables the arrhythmia analysis;
OFF: Disables the arrhythmia analysis.
PVCs ALM PVCs alarm
ON: When the PVCs alarm occurs, the monitor gives alarm
indications and saves the alarm;
OFF: When the PVCs alarm occurs, the monitor neither gives
alarm indications nor saves the alarm;
When OFF is selected, the
icon is displayed on the right of
PVCs in the parameter window.
ALM LEV Alarm level
Options: HIGH, MED and LOW.
ALM REC Alarm recording
ON: The monitor starts recording when the PVCs alarm occurs.
OFF: The monitor does not record when the PVCs alarm
occurs.
ALM HI Upper alarm limit
Determines the upper limit of the PVCs alarm. Range: 1–10.
An alarm is triggered when the PVCs exceeds the upper limit.
11.5.4 Arrhythimia Relearn
During ECG monitoring, you may need to start an arrhythmia relearn when a dramatic
change in the patient’s ECG pattern has occurred. A change in the ECG pattern could result
in:
Incorrect arrhythmia alarms,
Loss of ST measurements, and/or
Inaccurate heart rate.
Arrhythmia relearn allows the monitor to learn the new ECG pattern to correct arrhythmia
alarms and heart rate value, and to restore ST measurements. To manually start an arrhythmia
relearn, select the ARR RELEARN option in the ARR ANALYSIS menu. The “ARR
LEARNING” message will be displayed in the information area.
11-21