6.3 Alarm Statuses
When an alarm occurs, normally the monitor gives indications in the modes mentioned
above as per the alarm level. If necessary, you can set the monitor to the following alarm
statuses.
..
Alarms Disabled
..
Alarms Paused
..
System Silenced
..
Alarms Silenced
6.3.1 Alarms Disabled
If the alarm switch of a parameter is turned off, the monitor does not generate alarms even if
the measured parameter value exceeds the alarm limit. This status is called Alarms
Disabled.
To disable the alarms of a parameter, you should open the setup menu of the parameter.
Take heart rate (HR) as an example.
1.
Rotate the control knob and highlight the ECG parameter label.
2.
Press the control knob. The ECG SETUP menu pops up.
3.
Rotate the control knob and highlight the field on the right of HR ALM.
4.
Press the control knob, and then select OFF from the pull-down list.
5.
The HR alarms are disabled. The
icon will be displayed on the right of the ECG
parameter label.
NOTE
When a new parameter module is installed or when a parameter module is
turned ON, all the parameter alarms and technical alarms related to this
module are disabled in the first 30-second operating time. The other module
alarms are not influenced.