PVC 2* The PVC arrhythmia event occurs to the
patient.
BIGEMINY 2* The bigeminy arrhythmia event occurs to the
patient.
TRIGEMINY 2* The trigeminy arrhythmia event occurs to the
patient.
TACHY 2* The patient is suffering from tachycardia.
BRADY 2* The patient is suffering from bradycardia.
PNC 2* No pacer signal is captured.
PNP 2* The pacemaker is not paced.
MISSED BEATS 2* The arrhythmia event of missed beats occurs
to the patient.
IRREG
RHYTHM
2* Irregular rhythm occurs to the patient.
VENT
RHYTHM
2* Ventricular rhythm occurs to the patient.
MULTIF. PVC 2* Multiform PVCs occur to the patient.
C.2 Technical Alarm Messages
Note: XX represents the parameter modules like ECG, NIBP and SpO2, or the parameters
being monitored like HR, PR and SpO2.
The A field indicates whether an alarm can be completely cleared; the B field indicates
whether only visual and audible indications of an alarm can be cleared; the “L” field
indicates the alarm level, and 1 means high, 2 means medium, 3 mean low, * means the
level is user-adjustable.
C.2.1 General Alarm Messages of Parameter Modules
Alarm message A B L Cause Measure
XX INIT ERR N Yes No 1 XX module initialization error N Restart the
monitor. If the
error remains,
contact our
company for
repair.
Note: N stands for the error number.
XX COMM
STOP
No No 1 Failure in communication between
XX module and the main board.
XX COMM ERR Yes No 1 Error in communication between XX
module and the main board.
XX ALM LMT
ERR
No No 1 The alarm limit of the XX parameter
is changed inadvertently.
If the error
remains, contact
our company for