START
Selecting this option starts a measurement.
•
STOP
•
CANCEL
•
REC
•
Scale Y
•
Scale X
•
EDIT
•
EXIT
Measurement Procedure
In the process of the CO measurement, if the blood
temperature cannot restore to its initial value for a long time
and the measurement cannot stop automatically, you may
select this option to stop the measurement. The measured
CO and CI will be displayed.
Selecting this option during a measurement cancels the
measurement. Selecting it after a measurement deletes the
measurement result.
This option allows you to print the measurement curve using
the recorder.
This option allows you to change the scale of Y-axis
(temperature).
Three levels: 0 - 0.5., 0 - 1.
and 0 - 2.0..
You can adjust the scale according to the temperature
difference. The smaller the scale is, the larger the curve will
be.
This option allows you to change the scale of X-axis (time).
Two levels: 0 - 30S and 0 - 60S.