3.4 Installation Method
3.4.1 Bracket Mounting
For details, please refer to the corresponding instructions for use of bracket mounting.
3.4.2 Connecting to AC Power Supply
1.
Use the original three-wire AC power cord.
2.
Connect the power cord to the receptacle for AC power cord on the rear panel of the
monitor.
3.
Connect the other end of the power cord to a compatible 3-prong hospital grade AC
power outlet.
The 3-prong power outlet must be ground. If it is doubted, contact related personnel of the
hospital.
WARNING
•
Do not use three-wire to two-wire adapter with this instrument.
•
To avoid unexpected power interruption, do no use power outlet with a
wall-mounted switch control.
3.4.3 Installing the Battery
If the monitor is to be powered by the internal battery, install the battery following the steps
as below:
1.
Slide the battery door toward the rear of the monitor to open it.
2.
Move the battery catch above to one side using one finger.
3.
Insert the battery into the battery slot per the “+” and “-” indications.
4.
Move the battery catch to another side, and then insert the other one in the same way, if
your monitor is equipped with two batteries.
5.
Release the battery catch, and it will fix the battery.
6.
Close the battery door.
3-4