C.3 Prompt Messages
Prompt messages Cause Measure
ECG1 SIGNAL
SATURATION
Signals of abrupt change interfere with
the ECG signal.
Check whether the electrodes
and leads are well connected.
ECG2 SIGNAL
SATURATION
SEARCH PULSE The SpO2 module is searching the
pulse.
Wait till the end of the
searching.
DOMINANT
RHYTHM
CHANGE
A change happens to the patient’s
dominant rhythm.
None.
Sidestream CO2 module
CO2 STANDBY The CO2 module enters the
energy-saving status when switching
from the Normal mode to the Standby
mode.
None
CO2 START UP The CO2 module is starting. Wait for the CO2 module to
finish the startup.
CO2 CALIBRATE
ZERO
The CO2 module is in the zeroing
status.
Wait for the CO2 module to
finish the zeroing.
CO2 WARM UP The CO2 module is warming itself up
after startup.
Wait for the CO2 module to
finish warming itself up.
Recorder
RECORDER
INITIALIZING
The recorder is in the initializing status. Wait for the recorder to finish
the initialization.
RECORDER BUSY The recorder is recording. Wait for the recorder to finish
the recording.