5.
ALL: The monitor displays both the HR and the PR values. PR value is displayed at the
right of SpO2 label or IBP label, SpO2 takes priority of IBP . The monitor will alarm for
both abnormal HR and PR. HR is given priority in determining the source of the beep
tone. If HR is not available, the sound will be from the PR.
•
HR CHANNEL CH1: The heart rate is determined by the first ECG
waveform;
CH2: The heart rate is determined by the second ECG
waveform;
AUTO: The monitor automatically selects an optimal
channel to calculate the heart rate.
•
LEAD TYPE Options: 3 LEADS and 5 LEADS;
•
SWEEP Options: 12.5, 25.0 and 50.0 mm/s;
•
ECG DISPLAY
When LEAD TYPE is set to 5 LEADS, you can select one of the following options:
1.
NORMAL DISPLAY: The monitor displays two ECG waveforms.
2.
MULTI-LEADS DISPLAY: The monitor displays six ECG waveforms, which occupies
the whole waveforms area.
3.
HALF-SCN MULTI-LEADS: The monitor displays six ECG waveforms, which only
occupies four waveform positions.
When LEAD TYPE is set to 3 LEADS, only NORMAL DISPLAY can be selected for the
ECG DISPLAY.
Other Setup
Select OTHER SETUP >> in the ECG SETUP menu. The following menu appears.
Figure 11-7 Other ECG Setup
11-10