Shenzhen PM-9000 Express Manual del operador Pagina 295

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 339
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 294
AwRR
2 to 60 BrPM ±1 BrPM
> 60 BrPM Not specified
Drift meet the requirement of accurancy in 6 hours
Alarm range
CO2:
AwRR:
0 to 30 % (0 to 228 mmHg)
2 to 100 BrPM
Apnea alarm delay AwRR: 20 to 40 s
Refreshing rate 1s
Calibration Yearly calibration requested.
Calibration stability After module being used for 12 consective months, the error is < 1%
Rise time (10 % to 90 %)
Sampling flow
120ml/min, using the
DRYLINE™ water trap
and DRYLINE™
sampling line (2.5m)
CO2 250 ms (fall time 200 ms)
N2O 250 ms
O2 600ms
HAL, ISO, SEV, DES 300 ms
ENF 350 ms
Delay time < 4s