11.4 ST Analysis
11.4.1 Overview
•
The function of ST analysis is optional.
•
In the factory configuration, ST analysis is disabled.
•
When turning ST ANALYSIS on, the monitor selects DIAGNOSTIC mode
automatically. You can set the monitor to MONITOR or SURGERY mode as required.
However, the ST numerics might be severely distorted in these modes.
•
With the ST analysis, the variance of the ST segment at the waveform tracks of the
selected lead can be measured. The ST measurement result is displayed numerically in
the ST1 and ST2 positions in the parameter window.
•
You can review the ST trend graph and trend data in the TREND GRAPH and the
TREND TABLE menus.
•
Measurement unit of the ST segment: mV (millivolt).
•
Measurement symbols of the ST segment: “+” means positive elevation, “-” means
negative elevation.
•
Measurement range of the ST segment: -2.0 mV to +2.0 mV.
11.4.2 3/5-Lead ST Analysis Menu
Select ST ANALYSIS >> in ECG SETUP menu. The following menu appears.
Figure 11-9 ST Analysis
11-15
Comentarios a estos manuales