C.2.4 TEMP Module Alarm Messages
Alarm message A B L Cause Measure
TEMP1 No Yes 3 The T1 sensor is not Check for correct
SENSOR OFF connected correctly to the connection of the T1
patient or the monitor. sensor.
TEMP2 No Yes 3 The T2 sensor is not Check for correct
SENSOR OFF connected correctly to the connection of the T2
patient or the monitor. sensor.
TEMP No No 1 Circuit fault of the Contact our company for
SELFTEST temperature channel. repair.
ERROR
TEMP No No 2 Error in temperature channel Restart the monitor. If the
CALIBRATION calibration. error remains, contact our
ERR company for repair.
C.2.5 NIBP Module Alarm Messages
Alarm message A B L Cause Measure
NIBP
SELFTEST ERR
Yes Yes 1 Error in NIBP initialization. Select RESET from the
NIBP SETUP menu. If
the error remains, contact
our company for repair.
LOOSE CUFF No Yes 3 The NIBP cuff is not properly
connected.
Check the patient’s
condition, and check if
the patient type is
correct. Replace with a
proper cuff and connect
it correctly. If the
problem still exists,
contact our company for
repair.
AIR LEAK No Yes 3 Leak in the airway.
PNEUMATIC
LEAK
No Yes 3
CUFF TYPE
ERR
No Yes 2 The cuff applied is not
appropriate to the patient type.
AIR PRESSURE
ERROR
No Yes 3 Failure occurs in the pulse
measurement. The monitor
cannot perform measurement,
analysis, or calculation.
WEAK SIGNAL No Yes 3
SIGNAL