4.2.5 Discharge Patient
To discharge the patient being monitored, please follow this procedure:
1.
Select DISCHARGE PATIENT in PATIENT SETUP menu.
2.
Select YES in the pop-up menu.
3.
Prompt information will be displayed on the central monitoring system if the monitor is
connected with it.
4.3 Configuration Setup
Select CONFIG >> in the SYSTEM MENU. The following menu appears.
Figure 4-5 Configuration menu
If you have changed some settings during monitoring a patient and the changed settings are
not proper, especially when a new patient is admitted, you can restore the factory
configuration.
Restoring Factory Configuration
1.
Rotate the control knob and select the desired configuration.
2.
Select EXIT, and a CONFIRM CONFIG dialog box pops up.
3.
Select YES to restore the monitor to the selected configuration, or select NO to cancel
the operation.
4-6