A.16 AG Specifications
Measurement technique Infrared absorption
Measurement mode Sidestream
Warm-up time
45 seconds (warming-up status)
10 minutes (ready-to-measure status)
Sampling flow 70, 90, 120 ml/minute (user-selectable)
Gas type CO2, N2O, O2 (optional), Des, Iso, Enf, Sev, Hal
Measurement range
CO2:
N2O:
Des:
Sev:
Enf, Iso, Hal:
O2:
AwRR:
0 to 30%
0 to 105%
0 to 30%
0 to 30%
0 to 30%
0 to 105%
2 to 100 BrPM
Resolution
CO2:
AwRR:
1 mmHg
1 BrPM
Precision Gas Range (%REL) Precision (%ABS)
CO2
0 to 1 ±0.1
1 to 5 ±0.2
5 to 7 ±0.3
7 to 10 ±0.5
> 10 Not specified
N2O
0 to 20 ±2
20 to 100 ±3
Des
0 to 1 ±0.15
1to 5 ±0.2
5 to 10 ±0.4
10 to 15 ±0.6
15 to 18 ±1
>18 Not specified
Sev
0 to 1 ±0.15
1 to 5 ±0.2
5 to 8 ±0.4
> 8 Not specified