Shenzhen PM-9000 Express Manual del operador Pagina 242

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 339
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 241
17.2.5 Maintenance and Cleaning
The sampling line of the sidestream CO2 module is disposable and cannot be disinfected
for reuse.
In case of an exception in the sampling system of the CO2 module, check for
entanglement of the sampling line. If the sampling line is not entangled, remove it from
the water trap. In this situation, if the screen displays a prompt information indicating
the CO2 sampling line is abnormal, the water trap must have been blocked, and you
must replace it with a new one. If no prompt information is given, the sampling line
must have been blocked, and you should replace with a new sampling line.
Routine calibration of the sidestream CO2 module is not required. But the calibration
must be performed every year, or when great inaccuracy of measurement is found.
17.3 Removing Exhaust Gases from the System
WARNING
Anesthetics: When using the Sidestream CO2 measurement on patients who are
receiving or have recently received anesthetics, connect the outlet to a
scavenging system, or to the anesthesia machine/ventilator, to avoid exposing
medical staff to anesthetics.
To remove the sample gas to a scavenging system, connect an exhaust tube to the gas outlet
connector of the module.
17-11
Vista de pagina 241
1 2 ... 237 238 239 240 241 242 243 244 245 246 247 ... 338 339

Comentarios a estos manuales

Sin comentarios