AG HARDWARE
ERROR
Yes No 2 AG module
hardware fault. The
module remains AG
monitoring.
Unplug and re-plug the
module and then restart
the monitor.
AG DATA LIMIT
ERROR
Yes No 2 AG module fault.
AG ZREF FAIL Yes No 3 Failure in AG
module zeroing.
AG CAL FAIL Yes No 1 Failure in AG
module calibration.
CO2 ACCURACY
UNSEPCIFIED
Yes No 3 The measurement is
out of specified
range at specified
accuracy.
Check that appropriate
gas is used and the
clinical environment is
stable. If the problem
persists, send for
authorized maintenance
personnel.
N2O ACCURACY
UNSEPCIFIED
Yes No 3
O2 ACCURACY
UNSEPCIFIED
Yes No 3
HAL ACCURACY
UNSEPCIFIED
Yes No 3
ENF ACCURACY
UNSEPCIFIED
Yes No 1
ISO ACCURACY
UNSEPCIFIED
Yes No 3
SEV ACCURACY
UNSEPCIFIED
Yes No 3
DES ACCURACY
UNSEPCIFIED
Yes No 3