Prompt Information Related to Calibration
Take channel 1 as an example. The monitor may give the following prompt information after
the calibration has been finished.
•
CH1 SUCCESSFUL CALIBRATE
Channel 1 works normally and can be applied to IBP monitoring.
•
CH1 SENSOR OFF, FAIL
Verify the transducer of channel 1 is connected to the monitor, and then perform the
calibration again. If the monitor continues to give this prompt information, contact our
Customer Service.
•
IN DEMO, FAIL
Make sure the monitor is not in DEMO mode, and perform the calibration again. If the
monitor continues to give this prompt information, contact our Customer Service.
•
PRESSURE OVER RANGE, FAIL
Verify the preset calibration value is reasonable, and perform the calibration again. If the
prompt information is still displayed, contact our Customer Service.
•
PULSATILE PRESSURE, FAIL
Verify the pressure value indicated by the sphygmomanometer keeps unchanged, and then
perform the calibration again. If the prompt information is still displayed, contact our
Customer Service.