Shenzhen PM-9000 Express Manual del operador Pagina 214

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 339
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 213
Prompt Information Related to Calibration
Take channel 1 as an example. The monitor may give the following prompt information after
the calibration has been finished.
CH1 SUCCESSFUL CALIBRATE
Channel 1 works normally and can be applied to IBP monitoring.
CH1 SENSOR OFF, FAIL
Verify the transducer of channel 1 is connected to the monitor, and then perform the
calibration again. If the monitor continues to give this prompt information, contact our
Customer Service.
IN DEMO, FAIL
Make sure the monitor is not in DEMO mode, and perform the calibration again. If the
monitor continues to give this prompt information, contact our Customer Service.
PRESSURE OVER RANGE, FAIL
Verify the preset calibration value is reasonable, and perform the calibration again. If the
prompt information is still displayed, contact our Customer Service.
PULSATILE PRESSURE, FAIL
Verify the pressure value indicated by the sphygmomanometer keeps unchanged, and then
perform the calibration again. If the prompt information is still displayed, contact our
Customer Service.
15-12
Vista de pagina 213
1 2 ... 209 210 211 212 213 214 215 216 217 218 219 ... 338 339

Comentarios a estos manuales

Sin comentarios