C.2.12 Recorder Module Alarm Messages
Alarm message A B L Cause Measure
RECORDER
INIT ERR N
Yes No 2 An error occurs during the
recorder initialization.
Contact the hospital’s
engineers or our customer
Service.
Note: N represents the error number.
RECORDER
SELFTEST
ERR
Yes No 2 An error might occur to
the RAM, ROM and CPU
watchdog.
Open the RECORD menu
and select the CLEAR REC
TASK option. If the problem
remains, contact our
company for repair.
RECORDER
VLT HIGH
No No 1 A problem occurs to the
system power.
If this alarm message is
given for many times,
contact our company for
repair.
RECORDER
VLT LOW
No No 1
RECORDER
HEAD HOT
No No 1 The thermal head of the
recorder is too hot.
Resume the recording till the
recorder cools down
completely. If the problem
still exists, contact our
company for repair.
REC HEAD IN
WRONG POS.
Yes Yes 3 The thermal head of the
recorder is in wrong
position.
Restore the control lever of
the recorder to its previous
position.