NOTE
•
The apnea alarm cannot be disabled.
•
When the CO2 monitoring is not required, it is recommended the water trap not
be connected and the work mode be set to STANDBY.
Other Setup
Selecting OTHER SETUP option opens the following menu.
Figure 17-4 CO2 Other Setup Menu
You can perform the following settings in this menu.
•
PUMP RATE Determines the sampling rate of the CO2 module pump.
•
N2O COMPEN N2O compensation
The presence of nitrous oxide causes the CO2 value to be
higher than the actual value. Use this option to compensate
for the presence of N2O.
•
O2 COMPEN O2 compensation
The presence of a large concentration of oxygen causes the
CO2 value to be lower than the actual value. Use this option
to compensate for the presence of O2.
•
Des COMPEN Desflurane compensation
The presence of desflurane causes the CO2 value to be
higher than the actual value. Use this option to compensate
for the presence of desflurane.