A.15 CO2 Specifications
Measurement mode Sidestream
Measurement technique Infrared absorption technique
Displayed parameter EtCO2, FiCO2, Respiration Rate
CO2 measurement range 0 to 99mmHg
Precision*
0 to 40 mmHg:
41 to 76 mmHg:
77 to 99 mmHg:
±2 mmHg
±5%
±10%
Resolution 1 mmHg
Drift meet the requirement of accurancy in 6 hours
Sample flow rate 70, 100 ml/min
Precision of deflation rate ±15% or 15 ml/min, whichever is great
Start-up time of CO2
module
< 1min, the module enters the warming up status after the startup.
One minute later, it enters the ready-to-measure status.
AwRR measurement range 0 to 120 BrPM
Precision
0 to 70 BrPM:
> 70 BrPM:
±2 BrPM
±5 BrPM
Response time
When measured with a watertrap and a 2.5 m-long sampling line:
<5.5 s @ 100 ml/min
<7 s @ 70 ml/min
Delay time
When measured with a watertrap and a 2.5m-long sampling line:
<5 s @ 100 ml/min
<6.5 s @ 70 ml/min
Apnea alarm delay AwRR: 10 to 40 s
* Conditions for measurements in typical precision:
The measurement is started after the preheating mode of the module;
Ambient pressure: 750 mmHg to 760 mmHg; room temperature: 22. to 28.;
The gas under test is dry, and the balance gas is N2;
The deflation rate is 100 ml/min, the respiration rate is no greater than 50 BrPM, with a
fluctuation less than ±3 BrPM, and the inhale interval/exhale interval is 1:2;
When the working temperature is from 15 to 25 degree, or from 50 to 55 degree, or when the
breath rate is greater than 50Brpm, the measurement precision should be ±4mmHg (0 to
40mmHg) or ±12% of the reading (41 to 99 mmHg)