SpO2 and PR alarm limits:
Parameter Max. upper limit Min. lower limit Step
SpO2 100 0 1
PR 250 20 1
The SpO2 and PR alarm limits in factory configuration:
Parameter Patient type Upper limit Lower limit
Feline 100 90
SpO2 Canine 100 90
Other animals 100 90
Feline 180 120
PR Canine 110 60
Other animals 180 120
WARNING
Setting the SpO2 upper alarm limit to 100% will disable the upper alarm limit.
Therefore, the upper alarm limit for oxygen saturation must be carefully
selected in accordance with the commonly accepted clinical practices.
SWEEP Waveform speed
Options: 12.5 and 25.0 mm/s.
PR SOUND Pulse volume
Range: 0–10. 0 indicates the volume is closed and 10
indicates the maximum volume.
AVG TIME Average time
Determines the average SpO2 calculation time.
Options: 4S, 8S and 16S.
NIBP SIMUL ON: When measuring NIBP and SpO2 at the same side, the
SpO2 alarm status remains unchanged until the NIBP
measurement is finished;
OFF: The function of NIBP SIMUL is disabled.
CONFIG >> You can select this option to access the SPO2 CONFIG
menu, in which you may select FACTORY CONFIG or
USER CONFIG. After finishing your selection and exiting
the menu, the system pops up a dialog box asking for your
confirmation.
12-14