Shenzhen PM-9000 Express Manual del operador Pagina 178

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 339
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 177
Saturation (%)
SpO2%
Temperature
PCO2
2,3-DPG
PH
Temperature
PCO2
2,3-DPG
Fetal Hb
Saturation (%)
SpO2%
Temperature
PCO2
2,3-DPG
PH
Temperature
PCO2
2,3-DPG
Fetal Hb
Figure 12-4 Oxyhemoglobin Dissociation Curve
12.3.2 Precautions
WARNING
Pulse oximeter can overestimate the SpO2 value in the presence of Hb-CO,
Met-Hb or dye dilution chemicals.
ES (Electrosurgery) equipment wire and SpO2 cable must not be tangled up.
Carefully route patient cabling to reduce the possibility of patient entanglement
or strangulation.
Do not use this instrument and the sensors during magnetic resonance imaging
(MRI). Induced current could potentially cause burns. The monitor may affect
the MRI image, and the MRI unit may affect the accuracy of the oximetry
measurements.
Do not use the supplied sterile SpO2 sensors if the packaging or the sensor is
damaged. Return them to the distributor or manufacturer.
Before performing the testing, check the sensor cable. After unplugging the
SpO2 sensor cable from the socket, the system shall display the prompt
information "SPO2 SENSOR OFF" and give the audible alarm.
12-10
Vista de pagina 177
1 2 ... 173 174 175 176 177 178 179 180 181 182 183 ... 338 339

Comentarios a estos manuales

Sin comentarios