ALM LEV Alarm level
Options: HIGH, MED and LOW.
•
AMP ADJUST
Amplitude adjustment
This option allows you to adjust the waveform amplitude.
Options: MANUAL and AUTO.
1.
MANUAL: If MANUAL is selected, you can select an IBP channel manually from the
options, including ART, PA, CVP, RAP, LAP, ICP, P1 and P2, and then adjust the IBP
scale in the SCALE ADJUST >> menu.
2.
AUTO: If AUTO is selected, the waveform labels of IBP channels become P1 and P2
and the IBP scale is adjusted by the system automatically.
•
SWEEP Waveform speed
Options: 12.5 and 25.0 mm/s.
•
PR SOUND
Pulse volume
Range: 0–10. 0 indicates the volume is closed and 10
indicates the maximum volume.
•
FILTER
This option allows you to select the filter method of the IBP
waveform. Three options are available:
NORMAL: 16Hz filtering;
SMOOTH: 8Hz filtering;
NO FILTER: Displays the original waveform.
The default is NO FILTER.
•
RESPONSE
Options: HIGH, MED and LOW.
The corresponding averaging time is about 1 s, 8 s and 12 s
respectively.
The IBP value displayed on the monitor screen is the average of data collected within a
specific time. The shorter the averaging time is, the quicker the patient monitor responds to
changes in the patient’s blood pressure. Contrarily, the longer the averaging time is, the
slower the patient monitor responds to changes in the patient’s blood pressure, but the
measurement accuracy will be improved. For critically ill patients, selecting shorter
averaging time will help understanding the patient’s state.
•
IBP1 UNIT