This window contains the following information:
1.
Time span (the start time and end time of alarm recall).
2.
Alarm event type.
3.
The alarming parameter, parameter value, alarm level and the alarm time.
4.
The alarm event number (format: NO: n of N). N indicates the amount of alarm events
and n indicates the sequence number of the currently displayed alarm event.
5.
Parameter values at the time of the alarm event.
6.
Two waveforms at the time of the alarm event. You can set the waveform length by
selecting from the ALM REC TIME options in the ALARM SETUP menu. Please
refer to 4.4.2 Alarm Setup.
1
2
3
4
5
6
•
UP-DOWN
The monitor is able to store a maximum of 70 alarm events. But only one alarm event can
be displayed in the ALARM RECALL window once. You can select the UP-DOWN option
and then rotate the control knob to view an earlier or later alarm event.
Figure 9-5 Alarm Event Recall