TABLE 2
Guidance and declaration — electromagnetic immunity
The equipment is intended for use in the electromagnetic environment specified below.
The customer or the user of the equipment should assure that it is used in such an
environment.
Immunity test IEC 60601 Test
level
Compliance
level
Electromagnetic environment
— guidance
Electrostatic ±6 kV contact ±6 kV contact Floors should be wood, concrete
Discharge (ESD) ±8 kV air ±8 kV air or ceramic tile. If floors are
IEC 61000-4-2 covered with synthetic material,
the relative humidity should be at
least 30%.
Electrical fast ±2 kV for power ±2 kV for power Mains power quality should be
Transient/burst supply lines supply lines that of a typical commercial or
IEC 61000-4-4 ±1 kV for
input/output lines
(>3m).
±1 kV for
input/output
lines (>3m)
hospital environment.
Surge IEC ±1 kV differential ±1 kV different Mains power quality should be
61000-4-5 mode
±2 kV common
mode
mode
±2 kV common
mode
that of a typical commercial or
hospital environment.
Voltage dips, <5% UT <5% UT Mains power quality should be
Short (>95% dip in UT) (>95% dip in UT) that of a typical commercial or
interruptions and for 0.5 cycle for 0.5 cycle hospital environment. If the user
voltage variation 40% UT 40% UT of our product requires continued
on power supply (60% dip in UT) (60% dip in UT) operation during power mains
input lines IEC for 5 cycle for 5 cycle interruptions, it is recommended
61000-4-11 70% UT 70% UT that our product be powered from
(30% dip in UT) (30% dip in UT) an uninterruptible power supply or
for 25 cycle for 25 cycle a battery.
<5% UT <5% UT
(>95% dip in UT) (>95% dip in UT)
for 5 sec for 5 sec
Power frequency 3 A/m 3 A/m Power frequency magnetic fields
(50/60 HZ) should be at levels characteristic