You can perform the following settings:
MONI NAME
DEPT.
BED NO
NET TYPE
LOCAL NET NO
NOTCH FREQ.
LANGUAGE
AUX OUTPUT
Monitor’s name.
The department where the monitor is located.
The bed number where the monitor is located.
Network type
Options: CMS and CMS+.
It indicates the bed number of a monitor in the monitoring
network. If the NET TYPE is CMS, the LOCAL NET NO can
be set between 1 and 64.
Options: 50Hz and 60Hz;
If NOTCH is turned ON, the monitor filters the ECG signals
with the selected notch frequency.
Select the required language of the displayed texts.
Three options are available:
1.
ANALOG OUT (analog output)
If this option is selected, the auxiliary output port will be able to output analog
signals, and you can perform the settings in the ANALOG menu. For details, see
4.4.6 Analog Output.
If this option is not selected, the analog output function will be disabled and the
ANALOG >> option in the SYSTEM SETUP menu will become inactive. In this
situation, you cannot set the information in the ANALOG menu.