Shenzhen PM-9000 Express Manual del operador Pagina 285

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 339
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 284
Pace pulse
Pulse indicator
Pace pulses meeting the following conditions are marked by the
PACE indicator.
Amplitude:
Width:
Rise time:
±4 to ±700 mV
0.1 to 2 ms
10 to 100 µs
Pulse rejection
The heart rate meter rejects all pulses meeting the following
conditions.
Amplitude:
Width:
Rise time:
Min. input slew rate:
±2 to ±700 mV
0.1 to 2 ms
10 to 100 µs
20 V/s RTI
ST segment measurement
Measurement range -2.0 to +2.0 mV
Precision
-0.8 to +0.8 mV: ±0.02 mV or ±10%, whichever is greater
Beyond this range: Undefined
Update period 10 s
Arrhythmia analysis
Type
ASYSTOLE, VFIB/VTAC, PVC, COUPLET, VT>2,
BIGEMINY, TRIGEMINY, R ON T, MISSED BEATS,
TACHY, BRADY, PNC and PNP
A.9 RESP Specifications
Measurement technique Thoracic impedance
Lead Optional: lead I and lead II; default lead II
Respiration excitation
waveform
< 300 µA, sinusoid, 62.8 kHz (±10%)
Respiration impedance test
range
0.3 to 3 .
Baseline impedance range 200 to 2500 O (using an ECG cable with 1kO resistance)
Differential input
impedance
> 2.5 M.
Linear Signal Range 3 O p-p minimum
Bandwidth 0.2 to 2 Hz (-3 dB)
Sweep speed 6.25 mm/s, 12.5 mm/s, 25 mm/s
Vista de pagina 284
1 2 ... 280 281 282 283 284 285 286 287 288 289 290 ... 338 339

Comentarios a estos manuales

Sin comentarios