11.6.3 Respiration Setup
Selecting the RESP label on the screen opens the following menu.
Figure 11-18 RESP SETUP menu
In this menu, you can perform the following settings.
ALM Alarm on/off
ON: When a respiration rate alarm occurs, the monitor gives
alarm indications and stores the alarm;
OFF: When a respiration rate alarm occurs, the monitor
neither gives alarm indications nor stores the alarm;
When OFF is selected, the
icon is displayed on the right
of the RESP label.
ALM LEV Alarm level
Options: HIGH, MED and LOW.
ALM REC Alarm recording
ON: When a respiration rate alarm occurs, the monitor
enables the recording;
OFF: When a respiration rate alarm occurs, the monitor does
not enable the recording
ALM HI Upper alarm limit
Determines the upper limit of respiration rate alarm.
ALM LO Lower alarm limit
Determines the lower limit of respiration rate alarm.
11-27