A.8 ECG Specifications
Lead type
3-lead (1 channel):
5-lead (2 channels):
I, II, III
I, II, III, aVR, aVL, aVF and V
Lead naming style AHA, EURO
Lead fault
The lead resistance is no greater than 51 kO
and it is in parallel
with a 0.047 µF capacitor, it will not cause a lead fault condition.
For 3/5-lead, differential offsets = ±300 mV, it will not cause a
lead fault condition.
Sensitivity selection
1.25 mm/mV (×0.125), 2.5 mm/mV (×0.25), 5 mm/mV (×0.5),
10 mm/mV (×1), 20 mm/mV (×2) and AUTO
Sweep speed 12.5 mm/s, 25 mm/s, 50 mm/s
Bandwidth (-3 dB)
Diagnostic mode:
Monitor mode:
Surgery mode:
0.05 to 100 Hz (812A module)
0.5 to 40 Hz
1 to 20 Hz
Common mode rejection
Diagnostic mode:
Monitor mode:
Surgery mode:
=90 dB
=105 dB
=105 dB
(The notch filter is turned off.)
50/60Hz Notch Filtering The monitor provides software filtering against the 50/60HZ
industrial frequency.
In monitor and surgery modes, the 50/60HZ filter will be turned
on automatically.
In diagnostic mode, the 50/60HZ filter will be turned off.
Input offset current =0.1µA (except currents to drive leads)
Differential input impedance = 5M.
Input signal range ±8mV (peak-to-peak value)
DC offset voltage ±300mV
Accuracy of input signal
reproduction
Methods A and D were used to establish overall system error and
frequency response.
Auxiliary current (Leads off
detection)
Active electrode: < 0.1 µA
Reference electrode: < 1 µA
Patient leakage current < 10uA