Sterilization
Chemical Solution Sterilization
After finishing the cleaning, select an effective sterilant for chemical solution sterilization of
the operating room equipment. Buffered glutaraldehyde (e.g. Cidex or Hospisept) is
recommended. Do not use quaternary cationic detergents such as zephiran chloride. If the
whole unit is to be sterilized, immerse the transducer but not the electrical connector into the
sterilant for the recommended sterilizing period. Ensure that the dome has been removed.
Then rinse all transducer parts except the electrical connector with sterilized water or saline.
The transducer must be thoroughly dried before storing.
Gas Sterilization
For more complete asepsis, use gas sterilization. The transducer should be completely dry
after cleaning. When ethylene oxide gas is used as the gas disinfectant, follow the operating
instructions provided by the manufacturer of the gas disinfectant.
NOTE
The disinfectant temperature must not exceed 70°C (158°F). Plastics in the
pressure transducer may deform or melt above this temperature.
15-14