1.
NIBP label: Selecting this label to access the NIBP SETUP menu.
2.
Time of last measurement.
3.
NIBP unit: mmHg or kPa.
4.
Prompt information area: Shows the NIBP measurement mode and other information.
5.
Systolic pressure value (NS)
6.
Mean pressure value (NM)
7.
Diastolic pressure value (ND)
If a set of measured results appears in grey, it indicates this measurement is performed at
least 1 hour ago.
13.2 Monitoring Procedure
WARNING
The animal on which the NIBP measurement is done must be no less than 2kg.
NEVER perform NIBP measurements on patients with sickle-cell disease or
under any condition in which the skin is damaged or expected to be damaged.
ALWAYS set the correct cuff type on the monitor before performing
measurements. Incorrect cuff type setting may cause a danger to the patient
because the blood pressure level of the large cuff is higher than that of the small
cuff.
To perform NIBP measurement on a patient, follow the procedure as below:
1.
Power on the monitor.
2.
Check the patient information area on the screen. If the cuff type is incorrect, select a
correct cuff type in NIBP SETUP menu.
3.
Plug the air hose in the NIBP cuff connector of the monitor.
4.
Apply a cuff of proper size to the appropriate site of the patient.
5.
Connect the cuff with the air hose.
6.
Press the NIBP key on the control panel to start the NIBP measurement.
13-2