13.4.1 Calibration
If you select the CALIBRATE option, the monitor starts the NIBP calibration and the
CALIBRATE option changes to STOP CAL. Selecting the option again stops the calibration.
Calibrate the cuff pressure reading with a calibrated reference manometer (or mercury
manometer) with accuracy higher than 1mmHg. To perform the calibration, follow the
procedure shown below:
1.
Remove the blood pressure cuff from the monitor and replace it with a rigid metal
container or vessel with a capacity of 500 ml ± 5%.
2.
Connect a calibrated reference manometer (with an error less than 1mmHg) and a ball
pump using “T” connectors as shown below.
3.
Select the CALIBRATE option.
4.
Inflate the metal container using the ball pump until the reference manometer reads 0,
then 50, and finally 200 mmHg.
5.
The difference between the indicated pressure of the reference manometer and the
indicated pressure of the monitor will not exceed 3 mmHg. Contact our Customer
Service if these values are not met.
Monitor
Reference
manometer
NIBP cuff Hose
connector
Ball pump
Metal vessel
Figure 13-5 NIBP Calibration
NOTE
•
The calibration of the NIBP measurement should be performed every two years
or performed according to the Hospital Procedure.