1.
The curve for one of the six measurements and the corresponding CO value.
2.
The average CO value.
3.
The average CI value.
4.
The function option in the window for CO editing.
Values of selected measurements can be averaged and stored in the CO field in the HEMOD
menu for Hemodynamic calculations.
When you access the editing window for the first time, all the curves and the CO values of
valid measurements are highlighted, indicating these values are to be averaged. You can
rotate the control knob and select the unqualified curve, and then press the control knob to
exclude the measurement from the calculation. The excluded curve becomes white.
NOTE
The excluded curve can be selected and counted in the calculation again.
16.2.2 Blood Temperature Monitoring
Blood temperature monitoring can function when C.O. measurement is not taken. The
blood temperature is measured by the thermistor situated in the distal end of the catheter
in the pulmonary artery (see Figure 16-5).
The blood temperature alarm is disabled during the blood temperature measurement.
When the measurement ends, the alarm is enabled automatically.
The currently measured blood temperature is displayed in the CO parameter window.
Pulmonary artery
Balloon
Thermistor
Right artia
Catheter
Right ventricle
Figure 16-5 Catheter Site
16-7